HA5000P Automatic Hematology Analyzer Blood Cell Counter
■ 8.0”LCD with high resolution
■ 3-part differentiation of WBC, 20 parameters+3 histograms
■ 2 counting mode: whole blood and prediluted
■ Double apertures for WBC and RBC/PLT
■ Throughput: 60 samples per hour
■ Patented aperture technology: Automatic clog-clearing with high pressure and lower rate of clog
■ Up to 16000 sample results with histograms can be stored
■ Automatic diluting, lyzing, mixing, rinsing
■ Automatically cleaning sampling probe,High Stability & Accuracy
■ Built-in thermal printer
■ Reagent alarm system (optional)
■ Method:Coulter Principle;Colorimetry hemoglobin determination
■ Sampler Volume:Predilute: 20μL
■ Whole blood :10μL
■ Intelligent Reagent Management Function:Reagent information and residual display
■ Carryover :WBC≤1.5%, RBC, HGB<1.0%, PLT≤ 2%
■ Input and Output:1 RS232, keyboard, Mouse;4 USB, LAN


■ Parameters
WBC, LY#, MID#, GR#, LY%, MID%, GR%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT,MPV, PDW, PCT, P-LCR and Histogram for WBC, RBC, PLT Coulter Principle
■ Method
Coulter Principle
Colorimetry hemoglobin determination
■ Performance
Parameter |
Linearity Range |
Precision (CV%) |
WBC(109/ L) |
0.0-99.9 |
2.5(7.0-15.0) |
RBC(1012/L) |
0.0-9.99 |
2.5(3.50-6.00) |
HGB(g/L) |
0-300 |
1.5(110-180) |
MCV(fL) |
40.0-150.0 |
0.5(80.0-110.0) |
PLT(109/L) |
0-999 |
5.0(200-500) |
■ Dilution Rations
Whole blood |
Prediluted |
|
WBC/HGB |
1:333 |
1:676 |
RBC/PLT |
1: 42143 |
1:85858 |
■ Menu
Review, Test, Maintenance, QC, Setup, Shutdown, Prime reagents, System diagnosis, Aperture rinse, Soak and rinse, Calibration Instrument config.
■ Sampler Volume
Predilute, 20μL
Whole blood, 15μL
■ Aperture Diameter
WBC, 80μm
RBC/PLT, 70μm
■ Carryover
WBC, RBC, HGB<0.5%, PLT< 1%
■ Input/Output
1 RS232, 1 keyboard
■ Printout
Built-in thermal printer (2 optional print modes)
■ Operating Environment
Temperature15℃-30℃
Humidity 30%-85%
■ Power Requirement
AC100-240V,50/60Hz
■ Dimension
440mm(H)×350mm(W)×440mm(D)
■ Net Weight
17kg